206
_____________________________________________________________________________________________
8.5 Output Terminals
______________________________________________________________________________________________
NOTE
Comparison time and cycle vary depending on the type of calculation and recording
length.
J U D G E _ O U T
O K _ V A L U E
N G _ V A L U E
T
B U SY
ST O R A G E
T R IG _O U T
Operation of
the 8730-10
8731-10
Stop
Wait for
trigger
Storing
waveform
Comparison
processing
Display processing
Measurement start
Measurement stop
Stop
Clears previous
comparison result
Outputs comparison result
Holds comparison result
Trigger activated
Less than 500 ms
Referring Data
When conducting value comparison on a 1 kHz sine wave for No.1: CH1
maximum value, No.2: CH1 minimum value, No.3: CH2 maximum value,
and No.4: CH2 minimum value with time axis range: 100
μ
s/DIV, recording
length: 10 DIV, display format: single screen.
Comparison time (duration of comparison): Approx. 70 ms
Comparison cycle (from trigger to next trigger): Approx. 220 ms
2. Trigger Mode: Single, Auto Save: OFF