5.5 Value Comparison Details
______________________________________________________________________________________________
5.5.1 Flow of Setting Value Comparison and Measurement
NOTE
When the "
Trigger Mode
" is set to "
Single
", measurement continues until the
stop conditions are met and ends as soon as these conditions are met. When the
"
Trigger Mode
" is set to "
Repeat
" or "
Auto
", measurement continues until the
STOP
key is pressed.
Enabling the auto save function saves the data when the stop conditions are met.
For information on saving value calculation results, see (6) in 4.2.9 and 5 in
6.4.11.
5.5 Value Comparison Details
This section explains the parameters set under the "
Value
" tag.
The value calculation result is checked to determine whether it is between
the upper and lower limits. This enables the detection of an abnormal
waveform that deviates from the set range. The comparison result (PASS
or FAIL) is output from an external control terminal. (For details, see
8.5.1)
Make setting separately for each channel (except on the 8730-10).
Up to four calculations (No.1 to No.4) may be set. Calculations are made
in order from No.1 to No.4.
Value calculations can also be made without comparison.
The results of value calculations can be saved (see 4.2.9
⑥
and 6.4.11 No.
5)
This section describes the minimum settings and operating procedure.
This function makes calculations on a waveform to be stored.
1. Value Comparison Setting
OFF: Does not conduct value comparison.
ON : Conducts value comparison. Value calculations without comparison
can also be made.
2. Define the types of value calculations
Set the type and comparison of each calculation (No.1 to No.4).
3. Specify the range of calculations
Select whether to calculate for the entire waveform or a specified range.
4. Set the stop conditions
Select whether to stop measurement with PASS or FAIL.
The stop conditions may be combined with those for waveform area
comparison.
5. Press the START key to start measurement. Calculations and value
comparison are conducted on the stored waveform
The comparison result is displayed on the screen and output from an
external control terminal.
The value calculation result is displayed at the bottom of the screen.