5.5 Value Comparison Details
______________________________________________________________________________________________
5.5.2 Value Comparison
Setting Screen:
Value
Setting
5.5.3 Value Calculations
FAIL comparison
PASS comparison
FAIL comparison
Upper limit
Lower limit
Upper limit
Lower limit
Setting Screen: Value
No.1 to No.4 (20 kinds of settings such as "Average" etc.)
Turns value comparison ON/OFF.
When this setting is ON, you can set the parameters for value comparison.
Even if this comparison is turned OFF, the 8731-10, 8731-10 retains the
settings so that you do not need to set them again each time you turn this
setting ON or OFF.
Setting Procedure
1. Press the
COMP SETTINGS
key, and
select "
Value
" tag.
2. Select a value on the menu, then press
the
ENTER
key.
3. Use the
CURSOR
key to select "
OFF
" or
"
ON
", then press the
ENTER
key.
OFF: Does not conduct value comparison.
ON : Conducts value comparison.
Calculations without comparison can
also be made.
The upper and lower limits are set for each calculation result. The
calculation result is FAIL if the result falls outside this range.
You can set up to four calculations (No.1 to No.4).
1. Calculation Type Setting
Setting Procedure
1. Press the
COMP SETTINGS
key, and
select "
Value
" tag
2. Select "
No.1
" to "
No. 4
" on the menu,
then press the
ENTER
key.
3. Move the cursor to "
Type
", then press
the
ENTER
key.
4. Select a item, then press the
ENTER
key.
(See the following page for details of
calculations.)