![Hioki 8730-10 Instruction Manual Download Page 261](http://html1.mh-extra.com/html/hioki/8730-10/8730-10_instruction-manual_2137220261.webp)
245
_____________________________________________________________________________________________
13.1 Error Messages
______________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Warning Massage
Action
Warning 001
There is no waveform data.
Load waveform data
Warning 002
Keys are locked. To unlock, press the
&
To unlock the keys, press the "
" and "
" keys
simultaneously
Warning 011
There is no PC card.
Insert a PC card
Warning 012
Illegal format.
The recording format is not compatible with the 8730-10,
8731-10
Warning 013
The file cannot be opened.
The file may be damaged
Warning 014
The directory is full.
There is a limit to the number of files and directories that
can be created in the root directory
Warning 015
Path name error.
Do not use more than 127 characters
Warning 016
Same name exists(FILE).
Change the file name
Warning 017
Disk full.
There is insufficient free space on the medium to save the
data. Delete existing files or use a new medium
Warning 018
Same name exists(DIR).
Change the directory name
Warning 019
The card cannot be detected
Check the PC card
Chapter 13
Appendix
13.1 Error Messages
"
Warning
" appears on the screen if an error occurs.
In case of such warning, the ERROR terminal outputs a low-level signal.
(See 8.5.5 for details.)
Warning messages
The "
Warning
" indication appears only one time when error was happen.
(The massage is disappeared after few seconds)
The "
Warning
" massage is disappeared by pressing any key.
If the "
beep sound
" item on the adjustment screen is set to "
ON
", then the
beeper sounds intermittently while the message is displayed.
"Warning" message and explanation