5.5 Value Comparison Details
______________________________________________________________________________________________
5.5.5 Stop Conditions of the Value Comparison
Stop Mode
Details
PASS
Stops measurement when value comparison result is PASS.
FAIL
Stops measurement when value comparison result is FAIL.
PASS&FAIL
Stops measurement when value comparison result is PASS or FAIL.
Setting Screen:
Value
Stop Mode (PASS, FAIL, PASS & FAIL)
NOTE
The 8730-10, 8731-10 may be set to halt measurement according to the
result of value comparison (PASS or FAIL). Enabling the auto save
function saves the data (see 4.2.9 for details).
When area comparison is also selected, measurement ends according to
the common stop conditions (see 5.2.1 for details).
Setting Procedure
1. Press the
COMP SETTINGS
key, and
select "
Value
" tag.
2
.
Select "
Stop Mode
" on the menu, then
press the
ENTER
key.
3. Select a condition, then press the
ENTER
key.
When the "
Trigger Mode
" is set to "
Single
",
measurement continues until the stop
conditions are met and ends as soon as these
conditions are met. When the "
Trigger
Mode
" is set to "
Repeat
" or "
Auto
",
measurement continues until the
STOP
key is
pressed.
When the "
Trigger Mode
" is set to "
Repeat
"
or "
Auto
", measurement continues until the
STOP
key is pressed or STOP signal input.