96
Chapter
5
•
•
•
•
•
•
Mode Parameter
The mode parameter selects between the single unit mode and multiple unit mode. If you
have multiple units hooked up to the display and disable split screen mode, you will get data
from only one of the units.
On the Web page
, this parameter is not configurable via the web interface.
On the HI 6500 series instrument display
, select the display menu and use the UP or
DOWN button to select the Split Screen submenu, then select the Mode menu. The current
mode will be displayed, press the ENTER or the RIGHT button if the mode needs to be
modified.
Num Display Parameter
The Num Display parameter sets the number of displays to be shown in the split screen
mode. If a number less than four is selected, any display segments above the Num Display
value will be blanked (set to the background color)
On the Web page
, this parameter is not configurable via the web interface.
On the HI 6500 series instrument display
, select the display menu and use the UP or
DOWN button to select the Split Screen submenu, then select the Num Display menu item.
The current Num Display value will be displayed, press the ENTER or the RIGHT button
if this value needs to be modified.
Discover Command
This command starts the automatic HI 6500 series instrument network discovery of up to
four instruments. Once discovered, the instruments would be available for display in the
split screen mode.
On the Web page
, this parameter is not configurable via the web interface.
On the HI 6500 series instrument display
, select the Display menu and use the UP
or DOWN button to select the Split Screen submenu, then select the Discover menu
item then press the ENTER button.
Filter Parameter Menu
There are two parameters in the Filter menu
•
NumAverages (Number of Averages)
•
WAVERSAVER
Only the HI 6500-XP and the HI 6510-XP units have additional options for
WAV, Motion Threshold and Variation Threshold.
NumAverages Parameter
The value you enter for NumAverages (the Number of Averages) sets the number of weight
readings used to compute a sliding average of displayed weight. This helps reduce the
effects of material impact and/or vibration if material does not enter or exit the scale evenly.
This setting helps the instrument ignore the effects of material impact and/or vibration.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......