72
Chapter
5
•
•
•
•
•
•
Gravity Parameter
Gravity Correction compensates for an object weighing less at the equator than at the North
or South Pole.
Before modifying this value, please read the C2 calibration procedure described in Chapter
6.
On the
Web page
, select the calibration menu and left click inside the Gravity text field and
enter the desired Gravity value (see table in Chapter 6).
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the DO C2 eCAL submenu, then select the Gravity menu item. The
current value will be displayed, press the ENTER or the RIGHT button if this value needs
to be modified.
NumDevices Parameter
The NumDevices is a read only parameter that indicates the number of C2 eCAL load cells
that are connected to the HI 6500. This value is automatically calculated when Hardy C2
eCAL load cell(s) are detected during a C2 eCAL Calibration.
On the
Web page
, select the calibration menu and the number of C2 eCAL load cells
detected will be shown in the NumDevices text field.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the C2 eCAL submenu, use the IP or DOWN button the select the
NumDevices menu item and the number of C2 eCAL load cells detected will be shown on
the display.
Ref Weight Parameter
Reference Weight is the total live load that is currently on the scale. The calibration process
uses a reference weight, which is normally zero (no weight on the scale), but can be any
known weight on the scale. With nothing on the scale, the Reference Weight is 0.00. With
5 lbs on the scale, the Reference Weight is 5.00 lbs.
On the
Web page
, select the calibration menu and left click inside the Ref Weight text field
and enter the desired reference weight value.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......