Instrument Configuration
85
•
•
•
•
•
•
On the
HI 6500 series instrument display
, select the communication menu and use
the UP or DOWN button to select the Modbus-RTU submenu, then select the Baud
Rate menu item. The current Baud Rate value will be displayed, press the ENTER or
the RIGHT button if the value needs to be modified.
Parity Parameter
The Parity parameter has three options, EVEN, ODD, or OFF.
On the
Web page
, select the communication menu and then select the Modbus RTU
submenu, and left click inside the Parity field and select even, odd or off using the
drop down menu.
On the
HI 6500 series instrument display
, select the communication menu and use
the UP or DOWN button to select the Modbus-RTU submenu, then select the Slave
Address menu item. The current Parity value will be displayed, press the ENTER or
the RIGHT button if the Parity needs to be modified.
Termination Parameter
The Termination Parameter has three options, DISABLE, ENABLE and LINK.
Step 1. On the
Web page.
select the Termination drop down and select the termination
condition. Link means that the termination is set on or off depending on whether
pin 2 of the connector is low (on) or high (off). There is an internal pull up on pin
2 that will set the termination off if nothing is connected to it.
Step 2. Select Save Parameter to save the entry.
Step 3. Click Home to return to the home page.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......