128
Chapter
7
•
•
•
•
•
•
Zero Operation
In GROSS mode, remove all weight from the scale and wait until the scale has settled.
A scale in motion will show the symbol “
~
“ in the lower right hand corner of the screen
to indicate the scale has not settled. A scale that will not settle may have the Motion
Tolerance parameter established during SETUP adjusted too small for the environment.
Press the ZERO button, a message “ZERO OK” will briefly be displayed and the scale will
show
000.00
, indicating the ZERO command was successful.
If “ZERO FAILED” is displayed, the Zero command was unsuccessful. Check the
Zero Tolerance parameter established in Set Up to determine if the zero amount has
exceeded the maximum cumulative value that can be zero’d off the scale.
NOTE
The amount of weight zeroed off each time is cumulative. The zero command will fail
if the current gross weight plus any previously zeroed amount exceeds the zero
tolerance established in the SET UP parameters.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......