Network Configuration
49
•
•
•
•
•
•
•
6: WEIGHT SAMPLE CMD:
Write a #6 to the command register to run the Weigh
Sample command to set up the calibration of the Counts (if enabled).
•
Status Error code 1 (motion)
•
Status error code 2 (A/D error)
•
0x64 (100 decimal): CAL LOW CMD
. Write a 0x64 hex to the command register to
perform the low step of a traditional calibration. The status register will read 0 if this
command succeeds:
•
Calibration_Fail 1
•
Calibration_Fail_Motion 3
•
Calibration_Fail_Adc_Error 4
•
During a Cal Low Command: Code 2 indicates "calibration in progress".
•
0x65 (101decimal): CAL HIGH CMD
. Write a 0x65 hex to the command register to
perform the high step of a traditional calibration.
•
Calibration_Fail 1
•
Calibration_Fail_Motion 3
•
Calibration_Fail_Adc_Error 4
•
HardcalFailCounts 8: not enough counts between hard cal hi and hard cal lo
•
During a Cal High Command: Code 2 indicates "calibration in progress".
•
0x66 (102 decimal): C2 CAL CMD
. Write a 0x66 hex to the command register to
perform a C2 calibration.
•
Calibration_Fail 1
•
Calibration_Fail_Motion 3
•
Calibration_Fail_Adc_Error 4
•
Calibration_Fail_Noc2 5
•
Calibration_Fail_C2capeq 6
•
Calibration_Fail_C2clones 7
•
During a C2 Cal Command: Code 2 indicates "calibration in progress".
•
0x1000 (4096 decimal): WRITE INTEGER CMD
. Set the value of an integer
parameter. Write 0x1000 in the command register 0, the parameter ID number in
registers 2, 3 and the desired value in registers 4, 5.
•
No Error Codes
•
0x1001 (4097 decimal): WRITE FLOAT CMD
. Set the value of a floating point
parameter. Write 0x1001 in the command register 0, the parameter ID number in
registers 2, 3 and the desired float value in registers 4, 5.
•
Failure 1
•
TooHigh -1 : value is above legal limit
•
TooLow -2 : value is below legal limit
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......