Network Configuration
53
•
•
•
•
•
•
Step 2. Select termination drop down and select termination condition. Termination
maybe be set to Enable, Disable or Link. Link means that the termination is set on
or off depending on whether pin 2 of the connector is low (on) or high (off). There
is an internal pull up on pin 2 that will set the termination off if nothing is
connected to it.
Step 3. Click Save Parameters to save the entry.
Step 4. You can also read the communication status of the instrument.
Step 5. Click Home to return to the Home Page.
Initialization Process
To be able to add an HI 6500 OR HI 6510 to a Profibus-DP network, you need a PC or a
PLC and software such as Siemens Step 7™, Simatic Manager or equivalent, that allows
the Profibus-DP PLC and the HI 6500 or HI 6510 to exchange data
Profibus-DP .GSD File
All devices connected to a Profibus-DP network require a *.gsd
file. The *.gsd file contains all the parameters including the baud rate, table formats and
necessary data required by the network PLC when an HI 6500 or HI 6510 is connected to
the network.
You must download the proper *.GSD file from the Hardy website.
Step 1. Navigate from the products section to the HI 6500 web page.
Step 2. Click on the Tab, Docs & Programs and scroll down to Documents and Programs.
Select the *.GSD option to download the file.
Regardless of Simatic Manager you use, you must go through these three steps:
Step 1. Connect the HI 6500 Series Instrument to the Profibus DP network and verify the
address. (See the Installation Section for Installation and Cabling Instructions.)
Step 2. Step 2. Connect the PC to the Trunk Line. Load the configuration properties to the
initialization software on the PC and transfer them to the PLC.
Step 3. Step 3. Install the *.gsd file.
Pre-Initialization Procedures
Step 1. Inspect the network cables and make sure that the cables have been installed
correctly and satisfy the Profibus-DP guidelines for the data transmission baud
rate(s) required. (See the Cabling Chapter/Profibus Installation in this manual for
Profibus-DP cable specifications and cabling guidelines.)
Step 2. Select the Node address for the HI 6500 or HI 6510. as shown above. This can be
done before or after Initialization.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......