Calibration
125
•
•
•
•
•
•
Step 6. To set the Span Weight, place a certified test weight, equal to the span weight
entered in step 5, on the scale. This value is the amount of test weight that you will
apply, typically 80-100% of the expected maximum product loading on the load
cells.
Step 7. Use the DOWN button to select Do Cal Hi, then press the ENTER button to set
the Cal High point and complete the calibration sequence. A "Cal Completed OK"
message appears briefly if the calibration is successful. An Error number appears
if the calibration fails. See the Error list in the Troubleshooting chapter for help
with error correction
Multi-Point Calibration
Multi-Point Calibration, (also called midpoint correction) is simply a calibration with 3-5
points.
Standard two point calibration assumes linearity (a straight line). However, calibration to
just 2 end points may not accommodate for non-linearity (a curved line) that may happen
in the middle of the load curve.
Multi-Point calibration lets users calibrate at three, four or five values on the curve. A
multi-point calibration means, in comparison to a two point calibration, that you can
calibrate non-linearity significantly more accurately
To achieve the best possible accuracy, the multi-point calibration should cover the entire
range of the desired measurement values (from zero to the highest possible value).
To access Multi-Point Calibration from the web browser, click on the Multi-Point
Calibration link at the bottom of the Calibration page.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......