Network Configuration
39
•
•
•
•
•
•
•
HardcalFailCounts 8: not enough counts between hard cal hi and hard cal lo
•
During a Cal High Command,:Code 2 indicates "calibration in progress".
•
0x66 (102 decimal): C2 CAL CMD
. Write a 0x66 hex to the command register to
perform a C2 calibration.
•
Calibration_Fail 1
•
Calibration_Fail_Motion 3
•
Calibration_Fail_Adc_Error 4
•
Calibration_Fail_Noc2 5
•
Calibration_Fail_C2capeq 6
•
Calibration_Fail_C2clones 7
•
During a C2 Cal Command: Code 2 indicates "calibration in progress".
•
0x1000 (4096 decimal): WRITE INTEGER CMD. Set the value of an integer
parameter. Write 0x1000 in the command register 0, the parameter ID number in
registers 2, 3 and the desired value in registers 4, 5.
• No Error Codes
•
0x1001 (4097 decimal): WRITE FLOAT CMD.
Set the value of a floating point
parameter. Write 0x1001 in the command register 0, the parameter ID number in
registers 2, 3 and the desired float value in registers 4, 5.
•
Failure 1
•
TooHigh -1 : value is above legal limit
•
TooLow -2 : value is below legal limit
•
ErrorNotFound 0x8000 : parameter id not found
Status Word Bits
•
Bit 0 = A/D converter error - bad input from the load cells.
•
Bit 1 = A/D converter failure - no output from the converter to the processor.
•
Bit 5 = Wrong Platform - Trying to load incorrect firmware into unit.
•
Bit 6 = Motion - indicates weight is in motion (changing).
•
Bit 10 = EEPROM Write error - problem writing to the non volatile memory in the unit.
•
Bit 14 = IT test in progress - currently running Integrated Technician test.
•
Bit 15 = Not Found error - Parameter ID not found for command being run.
•
Bits 28-31 = counter - The upper 4 bits are constantly incrementing. This constantly
changing value can be used as confirmation of communications.
Other bits not listed here are not used at this time.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......