70
Chapter
5
•
•
•
•
•
•
Calibration Menu
•
C2 eCAL
o Sensitivity
o Gravity
o Ref Wt
o Do C2 Calibration
o Cal Tol
o Num Dev
•
Hard Cal
o Cal Tol
o Sensitivity
o Cal Lo Wt
o Do Cal Lo
o Span Wt
o Do Cal Hi
•
Cal Date
•
Multi-Point Calibration
C2 (eCal) Commands and Parameters
Cal Tolerance Parameter
Sometimes, the contents of the vessel you are weighing are in motion. Cal Tolerance allows
you to set a value that determines the amount of motion that the system can tolerate and still
calibrate. In other words, the value you enter for Calibration Tolerance sets the amount of
deviation to allow during the calibration process. This value must be greater than or equal
to the base motion value and/or the Graduation Sizes.
The base motion value can be calculated by using the following formula:
Base Motion value = (Total Load Cell Capacity x 0.0003)
RANGE: .000001 - 999999 (default 10)
Another way to determine what this value should be is to multiply the fluctuation from peak
to peak and multiply this amount by two
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......