176
Chapter
9
•
•
•
•
•
•
•
Raw variation for each load cell and the junction box ref weight.
•
WAVERSAVER® variation for each load cell and the junction box ref weight.
•
Raw variation results (pass/fail) for each load cell and junction box ref weight.
•
WAVERSAVER® variation results (pass/fail) for each load cell and junction box ref
weight.
•
The Return to Zero (RTZ) results (pass/fail) for each load cell and ref weight as well as
the combined reading.
To run the test users need to know the parameter ID (a 32 bit hex value) for the two
commands, Do IT Cmd and Do IT Reduced (voltage) Cmd.
These parameter IDs are:
0x3197 Do IT Test Cmd
0x3198 Do IT Reduced Cmd
To run the test, the user would set these parameters to a value of 1. Once the Test has been
run, the user can read each parameter ID they wish for the return value of that parameter.
The data will remain in the unit until the next time the IT test is run, which then would
overwrite the previous data with new data.
The parameter IDs are:
•
0x3197 Do IT Test Cmd
•
0x3198 Do IT Reduced Cmd
NOTE
The number of sensors reported is from a list and will show 1 less than the user's input for
the number of sensors for the selected jbox.
0x498DJBOX 1 NUMBER SENSORS
0x498EJBOX 2 NUMBER SENSORS
0x4990 WEIGHT CHAN0
0x4991WEIGHT CHAN1
0x4992WEIGHT CHAN2
0x4993WEIGHT CHAN3
0x4998WEIGHT JBOX REF1
0x4994WEIGHT CHAN4
0x4995WEIGHT CHAN5
0x4996WEIGHT CHAN6
0x4997WEIGHT CHAN7
0x4999WEIGHT JBOX REF2
0x49A0MV/V CHAN0
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......