Instrument Configuration
111
•
•
•
•
•
•
On the
HI 6500 series instrument display
, select the Security menu and use the UP or
DOWN button to select the Read Only submenu. Enter the password and then change the
current password press the UP or DOWN button and enter the current password then arrow
to "password" and enter and use the arrow keys to enter new password and press the
ENTER button to save and exit this submenu; and return to the top level security menu.
Setup Parameter Menu
The Setup Menu consists of the following parameters:
•
Capacity
•
Decimal Point
•
Grads
•
Instrument ID
•
Motion Tolerance
•
Operator ID
•
Units
Capacity Parameter
Scale Capacity is the scale's nominal operating capacity (the total weight capacity of the
scale system). If this value is exceeded by six graduations, dashes appear on the front
display. Communications to and from optional devices are not affected.
If a Capacity value is entered which conflicts with either the decimal point value or the
graduation value, the decimal point and/or graduation values are automatically modified to
match the programmed capacity. It is therefore recommended that the Capacity parameter
is entered first before setting the Decimal Point and Grad parameters.
RANGE: .000001 - 999999 (Default 999999).
On the
Web page
, enter the capacity in the text box provided
On the
HI 6500 series instrument display
, select the Setup menu and use the UP or
DOWN button to select the Capacity item. The current value will be displayed, press the
ENTER or the RIGHT button if this value needs to be modified.
Decimal Point Parameter
Use the Decimal Point Parameter to set the resolution you want for the instrument. Here
you set the location of the decimal point for the weight resolution. The higher the number,
the farther to the left the decimal point moves and the higher the resolution of the scale.
Note that setting more decimal points does affect the overall accuracy of the instrument.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......