Step 8. This message appears if no C2 load points are connected to the HI 6500. Either
do a traditional calibration or connect C2 load sensors to the instrument.
C2 Cal Failed: no C2 sensors found
Step 9. Click Back to return to the Calibration page.
C2 Calibration from the Front Panel
The example settings below are for illustrations only. Your setting requirements may
vary significantly.
Step 1. From the Configuration menu, use the UP or DOWN button to select Calibration,
and press the ENTER button to display the Calibration menu. Select C2 (eCal),
and press the ENTER button to display the C2 Calibration Menu items.
Step 2. Use the UP or DOWN arrow to select the NumDevices menu item. The read-only
NumDevices value provides the number of C2 (or eCal) load cells on the
system detected by the HI 6500 series instrument. Verify that the displayed
number matches the actual number installed. If not, check to ensure that each
load cell/point cable connection is securely fastened and that each load
cell/point cable is not broken. Only Hardy Process Solutions load sensors are
C2 (or eCal) calibration capable
Step 3. When you finish updating the value press Enter to save the Reference Weight and
return to the C2 eCal menu. Use the UP or DOWN button to select the Reference
Weight. The current reference weight will show. Press the ENTER or RIGHT
button if modifications to this value need to be made.
Step 4. Select Gravity Correction to display the value and press the ENTER button if the
value needs to be modified.
Step 5. Use the UP or DOWN button to select Do C2 eCAL and press the ENTER button
to perform the C2 eCAL calibration..
Step 6. A "Cal Completed OK" briefly appears on the screen indicating the C2 calibration
was successful.
•
A "Cal Failed" message briefly appears with the error number if the C2
Calibration was not successful. Chapter 9, Troubleshooting lists corrective
actions.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......