Troubleshooting
151
•
•
•
•
•
•
mV
DC voltage signals are between 0-25 millivolts. Overloads and negative millivolt
readings are not shown as actual readings but 25.5 mVfor over voltage and 0.0 for negative
voltage. You will need to use a multimeter with a 200 or 300mVDC range to view the out
of range voltages.
RTZ (Return to Zero) Test
This check is to see if the scale returns to a zero reading condition when it is empty. Run
this test only when the scale is empty. When the scale is calibrated at the
ZERO
or
Reference zero, there is a period during which the
WEIGHT
readings are collected. After
a valid calibration has been completed, the test compares the original weight reading with
collected and current readings. If the difference is more than the combined Motion and the
Zero Tolerance parameters the test will fail.
IT Test
If your system has an Integrated Technician Summing Junction box, the IT test can identify
individual load cell problems up to a maximum of four load cell selections. The dwell
displays are live readings and only available if you have an IT summing junction card, with
a maximum of one JBox and 4 channels. From the front display, select Jbox1.
From the Diagnostics menu, select Integrated Technician.
Sensor Number
Indicates which sensor is under test. Using the up or down arrow selects
the target sensor to be tested.Then select the number of load cells to test. and press Do IT
Test. The Dwell test for pass or fail take 30 seconds to settle. Until it settles, it will always
show “fail” when the test is started. If the browser times out during the test, refresh your
browser until the results appear.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......