Instrument Configuration
107
•
•
•
•
•
•
The Calibration Lock Parameters
Calibration Lock Parameter
When the Calibration Lock is enabled, the ability to run calibration through the front panel
is disabled unless the correct calibration password is entered by the operator. If the
Calibration lock is disabled, then calibration through the front panel is enabled.
On the
Web page
, on the security web page, select the Calibration Lock parameter, change
the password or leave it at the default password and then select either Enable or Disable
from the drop down list.
On the
HI 6500 series instrument display
, select the security menu and use the UP or
DOWN button to select the Calibration Lock submenu, select the Calibration Lock
submenu, press enter. You will be prompted for the current password. Enter the correct
password to access the menu. Select the Lock menu selection, then select the Calibration
Lock parameter. The current state, Enabled or Disabled, will be displayed. To change the
state of this parameter, press the ENTER button.
Calibration Password Parameter
The Calibration Password is a four digit alphanumeric character sequence. If the sequence
is not entered correctly through the front panel display the menu leaves the Calibration
Lock submenu unchanged.
On the
Web page
, select the security menu. The Calibration Password will be displayed on
the security web page.To change the Calibration Password, left click inside the text field
and enter a new 4 character alphanumeric password or leave it at the default password. The
current state, Enabled or Disabled, will be displayed. To change the state of this parameter,
change the value on the drop down menu.
On the
HI 6500 series instrument display,
select the Security menu and use the UP or
DOWN button to select the Calibration Lock submenu. To change the current password,
arrow down to the password menu item and press enter. Now using the arrow keys to
enter the new password. Press enter to accept. Press left arrow key to exit the menu. To
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......