Instrument Configuration
97
•
•
•
•
•
•
Applications requiring very quick weight readings should reduce this setting to its
minimum. If the weight is unstable, increase the averages. The HI 6500 series instrument
does 110 updates per second, which translates to an update approximately every 9
milliseconds. If you average enough weight readings, the weight loss or gain remains
smooth. If you average the weight too much you can cause over filling.
Also see
WAVERSAVER for information on filtering unstable weight readings.
RANGE:1-255 (default 10)
On the
Web page
, select the Filter menu selection and enter the value for the number of
readings to factor into the average. Our example (for illustration only) has set 10 for
Number of Averages. Yours will vary depending on your application.
On the
HI 6500 series instrument display
, select the Filter menu and use the UP or
DOWN button to select the NumAverages menu item. The current value will be displayed,
press the ENTER or the RIGHT button if this value needs to be modified.
WAVERSAVER® Parameter
Typically, mechanical noise (from other machinery in a plant environment) is present in
forces larger than the weight variations you want to detect. WAVERSAVER® reduces the
effects of the vibratory forces that exist in all industrial weight control and measurement
applications so the device can better calculate the actual weight. WAVERSAVER enables
the Weight Processor to distinguish between actual weight data and mechanical noise, both
of which are typically conveyed in the load cell signal. WAVERSAVER can be configured
to ignore noise with frequencies as low as 0.25 Hz. One of three higher additional cut off
frequencies may be selected to provide a faster instrument response time. The function is
user selectable and can be turned off.
RANGE: 0.25 Hz, 0.5 Hz, 1.0 Hz, 3.50 Hz, 7.50 Hz, OFF (default 1.0 Hz)
Chapter 1 describes the benefits of using the WAVERSAVER feature.
On the
Web page
, from the pull-down list, select the FILTER menu and then select the
value for WAVERSAVER.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......