Calibration
121
•
•
•
•
•
•
C2 and eCAL Calibration from the Web Page
From the Home Page, click Configuration to display the Configuration menu; then
click Calibration to open the Calibration page.
Step 1. Check to make sure the number of load cells displayed on the webpage
corresponds correctly to the inputs on the instrument.
Step 2. In the “Ref Weight” text field, enter the reference weight for your application.
Step 3. An empty scale will use a 0.00 reference weight setting.
Step 4. (Optional) If you want to enter the Gravity Correction Factor, see the table
above, then enter the Correction Factor number you selected from the table in
the Gravity Correction field
Step 5. Click the “Do C2 Calibration” button
Step 6. Wait a few seconds and the results will appear. If the Calibration was successful
a Cal completed OK appears and gross weight will read your reference value.
However, if no load points are connected to the HI 6500 series the following
message appears. Check the cable and connectors of the load point(s) and re-
calibrate.
Step 7. This message appears if no load points are connected to the HI 6500. Check the
cable and connectors of the load point(s) and re-calibrate.
Cal failed: A to D converter error
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......