5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-29
5.3.5 System shut
down messages
The system shut down message displays in the format
“System
shutdown
error
xxyy”
and is intended to be a help for complex failure analysis. The message consist of a
4 digit code
“ xxyy
”, that represents the last information status prior to the
microprocessor
systems
turn
down.
•
The “xx” represents the last “safe” status of micro controller A
•
The “yy” represents the last “safe” status of micro controller B
Error code Alarming uC Set condition
Possible failure
Action
Improper connection of
internal SIO cable
Check cable from front panel
bracket to X 14 on B board and
replace, if necessary
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
Defective uC controller A or
Defective safety diode on PC
board A
Replace PC board A
11yy A
Timer
3
system
problem
Defective Fuse SI 21, SI 22
Check fuse SI 21 and SI 22
and replace, if necessary
13yy A
Timer
6
system
problem
Defective uC controller A
Replace PC board A
14yy
A
Front side routine
without function
Defective uC controller A
Replace PC board A
ESD or EMI problem
Check environment for ESD or
EMI source
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
15yy
A
SIO data non plausible
Defective uC controller A
Replace PC board A
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...