7 Repair procedures
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a
7-36
7.9.8 Replace gas inlet
filters
Note
Each of the gas inlets has a sinter filter to protect the pneumatic assembly for
foreign material. If the filter's resistance is increased due to debris, the system will
alarm with a message during system test to clean the filter element. To replace the
filter:
1.
Remove the gas inlet connector according to section 7.9.7.
2.
Remove and replace the gas filter from the gas connector.
Note that the filter for AIR and O2 are identical.