9 Schematics and Diagrams
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
9-13
Mains line supply
AC line voltage 115 V
AC line voltage 230 V
DC mains voltage
Power consumption
Breathing circuit
Total compliance
Insp. resistance
Total resistance
Resistance test flow
Total leakage
Nebulizer supply pressure
Nebulizer flow range
Nebulizer gas source
Service menu
Line voltage AC
Line voltage DC
Panel voltage
Valve voltage
(Back up) battery voltage
Controller A voltage
Controller B voltage
Heater power
Exp. temperature
Insp. temperature
Gas supply
AIR supply pressure
O2 supply pressure
AIR pressure reducer setting
O2 pressure reducer setting
AIR supply pressure low signal
O2 supply pressure low signal
AIR supply flow
O2 supply flow
AIR supply flow low signal
O2 supply flow low signal
AIR 2
nd
stage flow valve turn on
O2 2
nd
stage flow valve turn on
Accepted range
98.0 to 132 V
AC,
50 to 60 Hz
198 to 263 V
AC
, 50 to 60 Hz
22.5 to 28.0 V
DC
< 50 W
Accepted range
0.5 to 5.0 mL/mbar
< 6.0 mbar/L/s (SW rev. 2.21 +)
< 15 mbar /L/s (SW rev. 2.21 +)
60L/min in Adult mode, 30 L/min in Pediatric mode
< 1,000 mL/min
16 to 20 kPa, 22.5 to 28.2 PSI (at given flow range)
4.0 to 12.0 L/min
AIR at FiO2 < 60%, O2 at FiO2 >= 60 %
Accepted range
19.0 to 29.0 V
DC
22.5 to 28.0 V
DC
12.5 to 17.0 V
11.5 to 15.0 V
22.5 – 29.0 V
DC
4.50 to 5.50 V
DC
4.50 to 5.50 V
DC
0.0 to 12.0 W
5.0 to 48.0 °C
5.0 to 48.0 °C
Accepted range
280 to 600 kPa, 40 to 80 PSI
280 to 600 kPa, 40 to 80 PSI
2.7 to 2.8 kPa, 38.0 to 39.5 PSI
2.7 to 2.8 kPa, 38.0 to 39.5 PSI
1.6 to 2.2 kPa, 23 to 31 PSI
1.6 to 2.2 kPa, 23 to 31 PSI
Minimum 110 L/min
Minimum 110 L/min
106 to 108 L/min
106 to 108 L/min
42 to 45 L/min
42 to 45 L/min
9.8 Accepted ranges
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...