6 Maintenance
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
6-3
6.2.1. Every twelve (12)
months
Perform the user maintenance from the procedures and schedules in the
Centiva/5 Plus CC ventilator User Reference manual.
Perform the following every twelve months:
1.
Remove all user level breathing system assemblies and components from
the breathing system.
2.
Remove the expiration valve from the main engine.
3.
Open the main engine according to section 7.2.5.
4.
Inspect the Centiva/5 ICU Ventilator and the removed components for
distortion, deterioration and/or cracks.
5.
Inspect and replace, if necessary, the expiration valve components
according to section 7.9.2.
6.
Replace the O2 sensor, if necessary, according to section 7.9.5.
7.
Replace O-ring on NIST connector (if applicable) at supply hose for AIR and
O2.
8.
Perform AIR flow sensor calibration according to section 4.7.3.
9.
Perform O
2
flow sensor calibration according to section 4.7.4.
10.
Perform expiratory flow sensor calibration according to section 4.7.5.
11.
Perform Airway pressure sensor calibration according to section 4.7.6.
12.
Perform Overpressure relief valve calibration according to section 4.7.7.
13.
Perform Battery charging voltage calibration according to section 4.7.8.
14.
Perform Battery test according to section 4.8.
15.
Reassemble and close the main engine.
16.
Perform the Post-service checkout according to section 3.1.
6.2 Maintenance schedule
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...