10 Configuration and optionalization
TRM, Centiva/5 Plus, M1047448/CM-922026, rev.3.50.a of 11/05
10-4
SW 3.nn.p HW o/r
Centiva/5 +
Bi-Level VG
Newborn
Trend
Abort
Abort
Safe
Save
Serial number
Option key
171050101
123456
Option menu
Language
Unit Meas.
Unit Co2
Unit IBW
Patient type
Cross over weight
Start vent. Mode
English
cmH2O
mmHg
kg
35 kg
Bi-Level
Adult
Default settings
To now enter from service mode into the configuration mode push the following
sequence of buttons within a time frame of 1 second:
1.
”x 10”
2.
“OPEN”
3.
“ComWheel”.
Configuration mode now is activated and the system prompts with the following
screen:
I
O
1
2
3
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...