5 Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-16
Code Message
Type
Set condition
Action
Reset condition
15-4 Limited flow in
oxygen gas
supply. Check
gas supply for
sufficient flow
delivery”.
Repeat test by
pressing
<Enter>.
Advice O2 flow sensor detects a
max flow of < 110 L/Min,
when 120 L/Min flow is
tested
Check gas supply hose,
plug, line and filters for flow
restrictions.
O2 flow sensor
detects > 110 L/Min
at 120 L/Min test
Or
repeat test by
pushing <Enter>
16-1 AIR flow sensor
miscalibrated
or defective,
measured
offset out of
range. Repeat
test by pressing
<Enter>.
Advice AIR flow sensor is
miscalibrated
Or
AIR flow sensor is defective
Recalibrate AIR flow line
AIR flow sensor
signal within proper
offset range
Or
repeat test by
pushing <Enter>
16-2 AIR flow valves
do not open or
no signal from
AIR flow sensor.
Check AIR gas
supply
Repeat test by
pressing
<Enter>.
Advice Insufficient AIR supply
pressure
Check for proper AIR gas
supply
AIR supply pressure
exceeding 2 bar (28
PSI)
Or
repeat test by
pushing <Enter>
16-3 “AIR flow offset
valve does not
react. Repeat
test by pressing
<Enter>.
Advice AIR flow sensor offset valve
problem identified
Replace B board
AIR flow offset valve
operates properly Or
repeat test by
pushing <Enter>
17-1 AIR valves do
not properly
close. Check
AIR gas supply
or repeat test
by pressing
<Enter>.
Advice AIR valves do not properly
close
Or
AIR gas supply pressure too
high
Or
AIR valve seat dirty.
Check AIR valve and
replace pneumatic
assembly, if necessary
AIR valves operate
properly
Or
repeat test by
pushing <Enter>
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...