4 Test and calibration
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
4-12
The service mode allows for verification of the following internal values of the
system:
6.
Line voltage = main bus supply voltage
a.
Line voltage
AC
= AC line supply detected
b.
Line voltage
DC
= DC line supply detected
7.
Panel voltage = control panel supply voltage
8.
Valve voltage = valve supply voltage
9.
Battery voltage = internal back-up battery voltage
10.
Controller A/B = microprocessor A/B supply voltage
11.
Heater power = expiratory valve heater power dissipation
12.
Temper. I/E = Inspiratory / expiratory temperature.
The service mode allows for verification of proper operation of the expiratory valve
locking mechanism.
13.
Open = unlock the expiratory valve assembly
14.
Close = lock the expiratory valve assembly
15.
Drawer = detected status of locking mechanism
a.
Drawer
< - ->
= exp. valve assembly removed
b.
Drawer
OK
= exp. valve assembly position for lock in properly
detected
The service mode allows the following calibration to be initialized:
16.
O2 cal = O
2
sensor calibration.
The service mode displays the following elapsed times:
17.
Power-on time = total time, device was supplied with mains supply.
18.
Operation time = total time, device was in a ventilation mode.
The service mode displays the following PEEP valve data (Exp. valve):
19.
G = gain value
20.
GR = gain value corrected
21.
O = offset value
22.
OR = offset value corrected
23.
K = correction factor
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...