5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-21
Code Message
Type
Set condition
Action
Reset condition
24-1
Exp. flow sensor
value too high
with AIR flow.
Check oxygen
sensor
calibration.
Check
expiration valve
for resistance,
ensure valve
matches
selected patient
type or repeat
test by pressing
<Enter>.
Advice
Expiration flow sensor
value too high with AIR
flow.
Or flow screen is dirty or
blocked
Or oxygen value incorrect
Or blue Newborn valve
assembly inserted with
Adult/Pedia selection
Check proper expiration
valve type is inserted
(silver for adult/pediatric
and blue for newborn)
or remove expiration valve
from lock, reassemble and
reinsert prior to restart the
system test
or recalibrate O2 cell
or recalibrate Air flow line
or replace pneumatic
assembly
Expiration flow
sensor signal in
proper range Or
Proceed with
system test by
pushing <Enter>
24-2
Exp. flow sensor
value too low
with AIR flow.
Check oxygen
sensor
calibration.
Check
expiration valve
for resistance,
ensure valve
matches
selected patient
type or repeat
test by pressing
<Enter>.
Advice
Expiration flow sensor
value too low with AIR
flow.
Or
Oxygen value incorrect
Or
silver Adult/Pedia valve
assembly inserted with
Newborn selection
Check proper expiration
valve type is inserted
(silver for adult/pediatric
and blue for neonate)
or remove expiration valve
from lock, reassemble and
reinsert prior to restart the
system test
or recalibrate O2 cell
or recalibrate Air flow line
or replace pneumatic
assembly
Expiration flow
sensor signal in
proper range Or
Proceed with
system test by
pushing <Enter>
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...