4 Test and calibration
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
4-4
If a malfunction is
identified during this
system test, the
corresponding message
appears in the screen,
including the according
failure code. For details
of error messages refer
to section 5.3.2
“Messages during
system test”.
After having passed the
system test, the system
prompts the user to
open the y-piece, to
allow testing for
inspiratory resistance.
Disconnect the y-piece
from the sealer. The
system will recognize
opening the y- piece
automatically and then
proceeds with the test.
After calculating the
inspiratory resistance,
the Centiva/5 Plus CC
ventilator will display the
measured data in the
screen.
SW 3.nn.p HW o/r
System test
Centiva/5
Exp. flow sensor value too high with AIR flow
Check oxygen sensor calibration
Check expiration valve for resistance
Ensure valve matches selected patient type
Repeat test by pressing <Enter>
+
Bi-Level VG
Newborn
Trend
# 24_1
Enter
Bi-Level VG
Newborn
Trend
SW 3.nn.p HW 0/r
System test
Centiva/5 +
Open y- piece
Adult
Adult
SW 3.nn.p HW 0/r
System test
Centiva/5
System test OK. To start ventilation <Enter>
Compliance
Resistance
Leakage
Nebulizer flow
+
Bi-Level VG
Newborn
Trend
2.3
0.8
60
7.3
ml/cmH2O
cmH2O/L/s
ml/min
L/min
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...