5 Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-28
No. Message
Type
Set condition
Meaning
Reset condition
20.
I=xx s, P=yy %,
E=zz s
Advice VCV,
SIMV-VC,
Bi-Level
VG mode:
A soft key with a direct
influence to the
ventilator timing was
activated
Information about the
absolute timing:
I= xx s; inspiration time =
xx secs
P= yy %; inspiratory
pause = yy % of
inspiration time,
E= zz s; expiration time =
zz secs
Push ComWheel to
confirm displayed data
21.
I=xx s, E=zz s
Advice
Bi-Level mode:
A soft key with a direct
influence to the
ventilator timing was
activated
Information about the
absolute timing:
I= xx s; inspiration time =
xx secs
E= zz s; expiration time =
zz secs
Push ComWheel to
confirm displayed data
22.
I:E=xx:yy, E=zz s Advice
Bi-Level mode in
newborn mode:
A soft key with a direct
influence to the
ventilator timing was
activated
Information about the I:E
ratio as a result of Tisnp
and rate setting:
I:E= xx yy; inspiratory to
expiratory ratio
E= zz s expiration time =
zz secs
Push ComWheel to
confirm displayed data
23.
“Switch
off
device? Yes:
Enter, No: any
button”
Advice
System in Standby
Or system open ?
And On/Off button
activated for more than
3 secs
Software switch off of
Centiva /5 by activating
the <Enter> button
Push any button to return
to normal operation
24.
“Battery
not
fully charged.
Switch off:
<Alarm silence
button>”
Advice
System in Standby
Or system open ?
And On/Off button
activated for more than
3 secs
AND internal back up
battery not fully charged
Battery is not yet fully
charged and system
denies the demand for
turn off. If necessary
Software switch off of
Centiva /5 by activating
the <Alarm silence>
button
Push any button to return
to normal operation
25.
“System
error
#
”
System
shut
down
message
Any system error
detected during
permanent system
check
Refer to section 5.3.5
“System shut down
messages”
26.
“Cross limits ?
Confirm”
Advice Newborn
mode:
Cross limit for either
Pinsp, Tinsp, PEEP or PS
is reached.
The system prevents to
pass the cross limit
Accept to pass the cross
limits by pushing the
ComWheel
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...