10 Configuration and optionalization
TRM, Centiva/5 Plus, M1047448/CM-922026, rev.3.50.a of 11/05
10-9
Note
Note
Centiva/5 Plus allows for optionalizating several already on board available
software options by activating a particular option code.
This optionalizating is only possible for one particular serial number and for one
particular software version and for the desired options.
The option key consists of information about
1.
serial number of the device to optionalize
2.
software version this option is valid for
3.
options desired
When ordering an option note the Centiva/5 Plus’ serial number, displayed in the
optionalization mode.
Software options come with a serialized 8 digit option key. This option key is printed
on a small sticker. Place this sticker after optionalization onto the rear side of the
Control Panel of the Centiva/5 Plus, so that in case it needs to be re-optionalized,
the option key is available at the machine.
If several options have to optionalized only one option key is necessary and one
optionalization clears all desired options at once,
With the current software release 3.01.p the following options can be cleared for
operation:
1.
Newborn
This option allows to ventilate patients with a body weight from 2
to 6 kg.
This option requires a dedicated (blue) newborn expiration valve
assembly, that comes with the option.
2.
Bi-Level VG
This option is a lung protective, automatic compliance
compensating pressure controlled ventilation mode with a
volume guarantee.
3.
Trend and event
This option allows to store, to display and to download all
measured data, ventilator settings, alarm settings over a period
of 14 days and up to 1000 events (e.g. alarms system messages
etc).
This optionalization mode configuration is not accessible to the user.
It is only accessible for authorized service personal.
10.2 Centiva/5 Plus optionalization mode
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...