5 Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-30
Error code Alarming uC Set condition
Possible failure
Action
ESD or EMI problems
Check environment for ESD or
EMI source
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
16yy
A
SIO data not received
Defective uC controller A
Replace PC board A
17yy
A
SIO cannot send
Defective uC controller A
Replace PC board A
18yy
A
AD converter defective Defective uC controller A
Replace PC board A
Cable problem between PC
board A and DCU board
Check cable and plug
Defective uC controller A
Replace PC board A
Defective PC board A
Replace PC board A
19yy A
Improper
communication
between uC A and
DCU board
Defective DCU board
Replace DCU board
20yy to
23yy
A Safety
relevant
data
non plausible
Defective uC controller A
Replace PC board A
Fuse SI 3 defective
Check fuse SI 3 and replace, if
necessary
Defective PC board A
Replace PC board A
30yy
A
5 V bus A out of range
Defective uC controller A
Replace PC board A
Fuse SI 1 defective
Check fuse SI 1 and replace, if
necessary
Defective PC board A
Replace PC board A
33yy
A
Panel supply voltage
out of range
Defective safety diode on PC
board B
Replace PC board B
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...