5 Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-26
5.3.4 Messages during
operation
No. Message
Type
Set condition
Meaning
Reset condition
1.
Battery xx Min. Advice In all modes:
If line power supply fails
Or
If external DC power
supply fails
Or
If internal battery is
charging
System power supply
from internal back up
battery. Remaining
operating time on
battery is “xx Min”
In all modes:
If line power is
reconnected
Or
If external DC power
supply is reconnected
Or
If internal battery is fully
charged
2.
Tube: a, xx, yy% Advice In all modes:
If Airway resistance
compensation ARC is
used
ARC is active for
a type of tube with
xx mm diameter and
yy % compensation
In all modes:
If ARC is turned off
3.
No test
Advice In all modes:
If system test had not
been performed
System test to identify
breathing circuits
compliance, resistance
and leakage was not
performed. System uses
default data to
compensate.
Perform system test
4.
Only O2
Advice In all modes:
If during system test
single gas operation for
O2 only was confirmed
Only O2 gas supply with
setting FiO2 default to
100% in all modes
Perform system test with
AIR gas supply
5.
Only AIR
Advice In all modes:
If during system test
single gas operation for
AIR only was confirmed
Only AIR gas supply with
setting FiO2 default to
21% in all modes
Perform system test with
O2 gas supply
6.
Adult
Advice In
all
modes:
System is in “adult” mode
For all system settings
adult mode default
values are used
Select “pediatric” or
“newborn” to change
7.
Pedia
Advice In
all
modes:
System is in “pediatric”
mode
For all system settings
pediatric mode default
values are used
Select “adult” or
“newborn” to change
8.
Newborn
Advice In
all
modes:
System is in “newborn”
mode
For all system settings
newborn mode default
values are used
Select “adult” or
“pediatric” to change
9.
Standby
Advice System was moved into
standby mode
System starts By-flow
after 15 secs to identify
reconnection of a patient
(APD). All alarms are
silenced
Connect a patient to the y-
piece
Or activate the On/Off
Button for more than 3
secs to turn off the system
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...