5 Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-14
5.3.2.Messages during
system test
Code Message
Type
Set condition
Action
Reset condition
Complete
system set up,
seal Y-piece.
Start system
test by
pressing
<Enter>
Advice System
has
successfully
passed power up test and is
ready to start system test
Complete system set up
with breathing system and
airway management
components intended to
be used upon following
breathing mode. Seal y-
piece by plugging y-piece
onto the sealer at the
expiration valve block
Proceed with system
test by pushing
<Enter>
12-1 No
O2
gas
supply.
Check O2 gas
supply or for
single gas AIR-
supply press
<Enter>
Advice Insufficient
O2
supply
pressure
Check for proper O2 gas
supply and for clean inlet
gas filters.
O2 supply pressure
exceeding 2 bar (28
PSI)
Or
repeat test by
pushing <Enter>
13-1 No
AIR
gas
supply.
Check AIR gas
supply or for
single O2-
supply press
<Enter>
Advice
Insufficient AIR supply
pressure
Check for proper AIR gas
supply and for clean inlet
gas filters.
AIR supply pressure
exceeding 2 bar (28
PSI)
Or
repeat test by
pushing <Enter>
14-1 Oxygen
flow
sensor
miscalibrated
or defective,
measured
offset out of
range.
Repeat test by
pressing
<Enter>.
Advice
O2 flow sensor is
miscalibrated
Or
O2 flow sensor is defective
Recalibrate O2 flow line
O2 flow sensor signal
within proper offset
range
Or
repeat test by
pushing <Enter>
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...