7 Repair procedures
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a
7-47
7.9.19 Replace serial
data connector
Note
The serial data connector is located on the rear of the main engine. To replace the
serial data connector:
1.
Remove the sub chassis according to section 7.9.14.
2.
Use a 5 mm nut driver to remove the nuts from the serial data connector.
3.
Remove the serial data connector to the outside.
4.
Replace the serial data connector.
Upon reinstallation pay attention to have the small side of the D sub 9 facing down.