4 Test and calibration
TRM, Centiva/5 Plus, M1047448/ CM-922026, rev. 3.50.a of 11/05
4-17
Adult
Pediatric
Newborn
Enter
Align
touch screen
Calibrate
O2 sensor
Bypass
system test
Unlock
exp. valve
Body weigth 55 kg
Ensure silver exp. valve is in place
Start system test by pressing <Enter>
SW 3.nn.p HW o/r
System test
Centiva/5
Complete system set up, seal y-piece
+
Bi-Level VG
Newborn
Trend
Retain
trend data
09:36:53
2005-03-21
4.7.1 O
2
sensor
calibration
Special notice
Access in user system test
menu
Special notice
The Centiva/5 Plus CC ventilator performs an automatic calibration of the O
2
sensor during the O
2
calibration routine. This routine requires an O2 gas supply.
The calibration routine performs the 100% (gain) calibration only.
To ensure proper gas mixture at the O2 sensor, connect a tube to the inspiratory
connector.
There are two ways to access the O2 calibration routine:
•
In the user system test menu.
•
In the service mode.
The calibration routines are identical.
Turn ON the
Centiva/5 Plus and
wait for the power-up
test to finish. The
system test screen
appears.
Select “Calibrate O2 sensor” and confirm the made selection by
pressing the ComWheel.
Ensure that the UBD (use before date) of the O2 cell is at least 6 months.
Calibrate
O2 sensor
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...