5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-27
No. Message
Type
Set condition
Meaning
Reset condition
10.
Apnea xx
Caution No breathing action is
detected within the last
apnea time delay.
Patient has stopped
breathing for more than
the apnea delay time.
Elapsed time is xx secs
Perform breathing
activities and start
ventilation support
11.
Apnea vent.
Caution No breathing action is
detected within the last
apnea time delay.
Patient has stopped
breathing for more than
the apnea delay time.
System has started
apnea back up
ventilation mode
Activate any ventilation
mode in the vent. mode
menu
12.
Pre oxygen xx.
Advice
Automatic suction
routine ASR activated.
1
st
phase
Pre oyxgenation phase of
automatic suction
routine is in operation
Deactivate automatic
suction routine ASR
Or Disconnect patient
13.
Suction xx.
Advice
Automatic suction
routine ASR activated.
2
nd
. phase
Suction is in process for
at least xx
Deactivate automatic
suction routine ASR
Or reconnect patient
14.
Post oxygen xx. Advice
Automatic suction
routine ASR activated.
3
rd
phase
Post oxygenation phase
of automatic suction
routine is in operation for
at least xx .
Deactivate automatic
suction routine ASR
15.
Freeze
Advice Graphic
freeze
function
was activated
Graphics in screen are
frozen free zed in place
for better analysis.
Deactivate freeze function
16.
System open ? Advice
Pressure below PEEP
Or patient disconnected
Or system disconnected
System cannot detect
sufficient pressure
increase.
Reconnect patient Or
Activate the On/Off button
for more than 3 secs to
turn off the system
17.
Advice The
nebulizer
function
was activated
Nebulizer function is
active.
Turn OFF nebulizer
function.
18.
Advice
The external DC supply
was activated.
System is supplied by
external DC power
supply.
Turn OFF or disconnect
external power supply OR
turn on AC mains line
supply.
19.
Advice
The AC supply was
activated.
System is supplied by AC
line power supply.
Turn OFF or disconnect AC
line supply OR
turn on DC mains line
supply.
+
-
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...