5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-5
No. Alarm
Priority Set condition
Action
Reset condition
VCV, SIMV-VC, Bi-Level VG,
Bi-Level mode:
System open (see messages)
for longer than Apnea time
High priority alarm.
VCV, SIMV-VC, Bi-
Level VG, Bi-Level
mode:
Next expiration is
detected and
Rate > 3 bpm
4 “Apnea”
High
CPAP/PSV mode:
Since the last expiration no
breath is detected in set
apnea time
or measured rate < 4 bpm
High priority alarm.
Bi-Level mode is started.
Apnea time displayed on
screen.
After 5 min flashing
visual alarm and audible
alarms are turned OFF
and message remains in
screen.
CPAP/PSV mode:
Restart CPAP/PSV
mode or start any
other mode
and
next expiration is
detected
and
Rate > 3 bpm.
5 “Leakage” High In
all
modes:
If ((V
insp
-V
exp
) / V
insp
) * 100 >
set leakage (5% - 95%)
over a rolling average of 4
breaths
High priority alarm.
In all modes:
If ((Vi
inspi
-V
exp
) / V
insp
) *
100
≤
set leakage (5%
- 95%) within one
breath
6 “Vt
not
delivered”
High
VCV, SIMV-VC, Bi-Level VG
mode:
If P
peak
≥
P
limit
And
Vt measured
≤
Vt set * 0.8
after 15 sec from start of
inspiration
High priority alarm
VCV, SIMV-VC, Bi-
Level VG mode:
If P
peak
< P
limit
And
Vt measured > Vt set x
0.8 at the end of
inspiratory time
7 “MV
low” High In
all
modes:
If calculated MV over 61
seconds or over 10 breaths
(which ever is smaller) < set
MV limit (0.5 – 50 L/min)
High priority alarm.
In all modes:
If calculated MV > set
MV limit (0.5 – 50
L/min)
8
“MV high”
High
In all modes:
If calculated MV over 61
seconds or over 10 breaths
(which ever is greater) > set
MV limit (0.5 – 50 L/min)
High priority alarm.
In all modes:
If calculated MV < set
MV limit (0.5 – 50
L/min)
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...