7 Repair procedures
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a
7-45
7.9.17 Replace DC
mains connector
The DC mains connector is located on the rear of the main engine. To replace the
DC mains connector:
1.
Remove the sub chassis according to section 7.9.14.
2.
Disconnect the wires from connector X 22.
3.
Use a 17 mm wrench to remove the nut from the connector on the inside of the
housing.
4.
Remove the DC mains connector to the outside.
5.
Replace the DC mains connector.
6.
When replacing the DC mains connector be sure to reconnect the wires to
connector X 22 in the proper order.
X 22
orange
blue
red
4
5
6
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...