4 Test and calibration
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
4-26
Note
Continuation from previous page
25.
Check the “Exp. flow” “Actual” to read “Cal10”
±
0.5 L/Min.
26.
If not, go back to step 18. Change setting for 30 l/min within the tolerance in
the desired direction to match the “Cal10”
±
0.5 L/Min.
27.
Use the ComWheel to increase the AIR flow to read 60 L/min on the “AIR
flow” “Rated” line.
28.
Check the “Exp. flow” “Actual” to read “Cal60”
±
3.0 L/Min.
29.
If not, go back to step 18. Change setting for 30 l/min within the tolerance in
the desired direction to match the “Cal60”
±
3.0 L/Min.
30.
When all three values match, calibration is performed properly.
31.
Now unscrew flow screen
retainer in the exp. valve
assembly. For flow retainer with
a visible slot use the stainless
steel plate of the PEEP valve to
unscrew. For other flow retainer
use the inspiration connection
nut from the tool kit ST4.
32.
Remove “Service” Flow screen
and insert the original flow
screen (customer flow screen).
33.
Screw the flow retainer.
34.
Perform a system test to verify proper operation of the valve assembly.
For Centiva/5 NB (newborn extension) use always the silver Adult / Pediatric
(standard) expiration valve with the according “Service “ flow screen to perform
expiration flow sensor calibration
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...