5 Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-8
No. Alarm
Priority Set condition
Action
Reset condition
17
“Rate low”
High
In all modes:
If measured rate < set rate
low limit
High priority alarm
In all modes:
If measured rate
≥
set
rate low limit
18 “Rate
high” High In
all
modes:
If measured rate > set rate
high limit
High priority alarm
In all modes:
If measured rate
≤
set
rate high limit
19 “Power
PCB
fail”
High In
all
modes:
If main bus voltages are
failing
High priority alarm
In all modes:
If main bus 13v
≤
15 V
And
If main bus 15v
≥
13 V
And
If main bus 15v
≤
17 V
20 “Display
failure”
High In
all
modes:
If connection between
control panel and main
engine is defective
Or if display back light is
defective
High priority alarm
In all modes:
If CPU handshake is
OK
And
If display back light
supply
≥
12 V
21
“O2 sensor fail” High
In all modes:
If O2 measured value
≤
10 %
Or
If O2 measured value
≥
120
%
High priority alarm
In all modes:
If 10% < O2 measured
value < 120 %
22 “O2
concentr.
low”
High In
all
modes:
If O2 measured value < set
O2 low limit
High priority alarm
In all modes:
If O2 measured value
≥
set O2 low limit
23 “O2
concentr.
high”
High In
all
modes:
If O2 measured value > set
O2 high limit
High priority alarm
In all modes:
If O2 measured value
≤
set O2 high limit
24 “P
limit
reached”
Low
VCV, SIMV-VC, Bi-Level VG
mode:
If Ppeak
≥
Plimit
Low priority alarm
VCV, SIMV-VC, Bi-Level
VG modes:
If Ppeak < Plimit at the
end of inspiratory time
25
“Button failure” Low
In all modes:
If a button is active for more
than 5 sec
Low priority alarm
Release button or
replace front label
26
“DC supply fail” High
In all modes:
If external 24 V DC supply
fails or
If external battery pack
voltage < 22.5 V
High priority alarm
Restore 24 V DC
supply or
restore external battery
pack with fully charged
one or
supply system by AC
line supply
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...