8 Illustrated parts
TRM, Centiva/5 Plus, M1047448/ CM-922026, rev. 3.50.a
8-3
8.2.4 Tool kit ST 4
8.2.5 Test equipment
specifications
The tool kit (P/N MA-640007) consist of
•
Pozidrive, screw driver, PH L1
•
Inspiration connection nut
•
Nut driver, size M 7
•
Allen wrench, size M 3
•
Nut driver, size M 5.5
•
Nut driver, size M 20/22
•
Tightening key
•
Calibration screw driver
•
EPROM grip
•
Tool bag
This tool kit is designed for field service. Other tools can be used as long as their
sizes match the above.
Instrumentation used to service the Centiva/5 ICU Ventilator must meet or exceed
the following specifications:
•
Pressure meter
•
Pressure range
0 – 90 mbar
•
Accuracy
±
0.1 mbar
•
Flow meter
•
Flow range
0 – 150 L/Min
•
Accuracy
±
0.05 L/Min or 2% of reading,
•
Calibrated for
AIR and O2
•
Multi meter
•
Type
Standard multi meter
•
Parameter
V
AC
, V
DC
, A
AC
, A
DC
, Ohm
•
Accuracy
±
1%
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...