4 Test and calibration
TRM, Centiva/5 Plus, M1047448/ CM-922026, rev. 3.50.a of 11/05
4-35
Panel voltage
Valve voltage
Battery voltage
Controll. A/B
Heater power
Temperature
Drawer
ExpValve
The panel voltage measures the +13/+15V control panel supply. Accepted voltage
range is 12.5 to 17.0 V.
The valve voltage is measured at the main bus 13 V WD (WD = watchdog controlled
bus). Accepted voltage range is 11.5 to 15.0 V.
The battery voltage measures the internal back-up battery voltage. Accepted
voltage range is 22.5 to 29 V. Additional data:
•
U
batt
> 29.0 V = Battery failure alarm
•
U
batt
< 22.5 V = Battery low alarm+
•
U
batt
< 22.0 V = Battery depleted alarm
The controller voltage is measured for each microprocessor, the micro A and micro
B supply bus. Accepted voltage range is 4.5 to 5.5 V.
The heater power measures the power dissipation of the expiratory flow sensor
heating system. Accepted power dissipation is < 12.0 W.
The temperature measures the expiratory flow sensor temperature. Accepted
temperature is 5 to 48 °C.
“OK” indicates proper locking of the expiratory valve.
“< -- >” indicates properly unlocking of the expiratory valve.
These data are used for manufacturing QM only.
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...