5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-15
Code Message
Type
Set condition
Action
Reset condition
14-2 Oxygen
flow
valves do not
open or no
signal from
oxygen flow
sensor.
Check O2 gas
supply and
press
<Enter>.”
Advice Insufficient
O2
supply
pressure
Check for proper O2 gas
supply
O2 supply pressure
exceeding 2 bar (28
PSI)
Or
repeat test by
pushing <Enter>
14-3 Oxygen
flow
offset valve
does not react.
Repeat test by
pressing
<Enter>.
Advice
O2 flow sensor offset valve
problem identified
Replace B board
O2 flow offset valve
operates properly
Or
repeat test by
pushing <Enter>
15-1 Oxygen
valves
do not properly
close.
Check O2 gas
supply or repeat
test by pressing
<Enter>.
Advice
O2 valves do not properly
close
Or
O2 gas supply pressure
too high
Or O2 valve seat dirty.
Check O2 valve and
replace pneumatic
assembly, if necessary
O2 valves operate
properly
Or
repeat test by
pushing <Enter>
15-2 2
nd
oxygen flow
valve does not
react. Repeat
test by pressing
<Enter>.
Advice
O2 flow valve problem
identified
Check O2 valve and
replace pneumatic
assembly, if necessary
2
nd
O2 valve
operates properly
or
Proceed with system
test by pushing
<Enter>
15-3 Malfunction
in
oxygen valve
testing. Test
takes too long,
repeat test by
pressing
<Enter>.
Advice
O2 flow valve problem
identified
Check O2 valve and
replace pneumatic
assembly, if necessary
Repeat test by
pushing <Enter>
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...