7 Repair procedures
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a
7-30
7.9.2 Repair expiratory
valve
The Centiva/5 Plus CC Ventilator is equipped with an expiration valve locking
mechanism that uses a motor driven locking system to properly fix the expiration
valve inside the system. To repair and inspect expiration valve:
1.
Remove the expiration valve according to section 7.2.1.
2.
Remove membrane plate (1) to gain access to the PEEP membrane
assembly (2).
3.
Remove the PEEP membrane assembly (2) from the base of valve (3) and
replace membrane, if necessary.
4.
In case the PEEP membrane assembly (2) is apart, lock the PEEP membrane
assembly together by inserting the silicon lock at the PEEP top seal
membrane (2a) through the PEEP PSU plate (2b) (cut edge facing 2c) and
then through the PEEP silicone membrane (2c). Pull on the lock until the
locking lip is fully engaged.
5.
Insert the PEEP membrane assembly into the expiration valve with the
sealing membrane (2c) facing down. The side of the top seal (2a) marked
“top” should be facing the top.
6.
Inspect and clean flow measurement port.
7.
Replace and grease o-ring (5) at flow measurement port.
8.
Use the inspiration connection nut of service tools to remove the flow screen
retainer (8) by turning counterclockwise.
9.
Carefully remove the flow screen (7) and o-ring (6).
10.
Insert o-ring (8) and the flow screen (7).
11.
Use the inspiration connection nut of service tools to fix the flow screen
retainer (8) by turning clockwise.
12.
Verify proper function after reassembly by performing a system test.
Summary of Contents for Centiva/5 Plus
Page 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Page 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Page 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Page 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Page 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Page 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Page 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Page 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Page 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Page 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Page 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...