System Overview
2-27
UDB_OVERCUR_N
Over current error
Digital
0-+15v
J1-C9
0V= effective
DOWN_PROTECT_TO_MCB_N
Reach limit
Digital
0-+15v
J1-C10
0V= effective
UP_TO_MCB_N
UP key
Digital
0-+15v
J1-C8
0V= effective
DOWN_TO_MCB_N
DOWN key
Digital
0-+15v
J1-C20
0V= effective
DOWN_SECURITY_TO_MCB_N
DOWN_SECURITY key
Digital
0-+15v
J1-C21
0V= effective
Workstation Communication
Signals from MCB to WK (Workstation) are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
WK_Y
One of two differential signals from MCB to WK
Differential
signal
+/-(1.5-
5V)
J2-B2
-
WK_Z
The other one of two differential signals from
MCB to WK
J2-A2
-
Signals from WK to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
WK_A One of two differential signals from WK to MCB
Differential
signal
+/-(1.5-
5V)
J2-A3
-
WK_B The other one of two differential signals from WK
to MCB
J2-B3
-
Display Board
Signals from MCB to Control Panel are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
SPI_CLK
Clock of SPI bus
Digital
0/+15V
J1-B23
-
SPI_MOSI
Data from MCB to Control Panel
Digital
0/+15V
J1-C24
-
SPI_/CS1
Chip select 1 for MAX9654
Digital
0/+15V
J1-A25
+15V= effective
SPI_/CS2
Chip select 2 for MAX7301
Digital
0/+15V
J1-A28
+15V= effective
LED_
YELLOW_N
Yellow led lights signal
Digital
0/+15V
J1-C23
0V= effective
Signals from Control panel to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
SPI_MISO
Data from Control Panel to MCB
Digital
0/+15V
J1-A23
-
DB_INT
Interrupt of Control panel
Digital
0/+15V
J1-B25
+15V= effective
User Interface Control
Key Switch
Signals from key switch to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
KEYSW_ON
The termination which
indicates lift column and x-ray
are both enable
Passive
Open/Close
J1-C26 KEYSW_ON & KEYSW_COM is
connected means both lift column
and exposure enable function
have been selected.
KEYSW_OFF The termination which
indicates lift column is enable
and x-ray is disable
Passive
Open/Close
J1-B16 KEYSW_OFF & KEYSW_COM is
connected means lift column
function have been selected.
KEYSW_COM Common termination
Passive
-
J1-A24 -
Hand Switch
Signals from hand switch to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
HAND_FLU
Hand switch
termination
Passive
Open/Close J1-B28 HAND_FLU & HAND_COM is
connected means HAND_FLU
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...