Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-18
If the footswitch or hand switch is released before completion, you should press the
switch again until the control panel LED displays “END END”.
If the “Er* Er*” is displayed on the control panel, you should redo the mA calibration
procedure from step 3.
16.
Make sure the footswitch or hand switch is released. Remove the jumper on J5 and reboot the
system. Take exposure at 80kV, 0.1mA. Check the measured current value should be 0.20
±
0.01mA. Otherwise, adjust POT4 on MCB board. Clockwise rotation will increase the current. And
anticlockwise rotation will decrease the current.
17.
Take exposure at 80kV, 2mA. Measure TP7 (MA-SMP) on MCB board with oscilloscope. Set
oscilloscope in TRIGGER mode and the trigger level is –250mV. The first pulse wave should be
orthogonal and the mA feedback curve shall be converged within 10ms, similar to the picture
below:
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...