Replacement
7-35
Warning 0117800 may occur when Brivo OEC 715/785/865 system boot up with a new
MCB installed. Do not shut down the system by pressing the confirm button. Instead, hold
down Press keys “Ctrl +Shift+Alt+
→
” to invoke the service tool first in Administration
interface. Then, refer to the chapter 4 and follow the instructions to finish the calibrate
process.
For MCB with Firmware 1.9.0-F or lower version, when replacing new version MCB, WKS
APP SW should be also upgraded to WKS 1.0.
Filament Drive Board (P/N: 5085604)
1.
Remove the rear cover of the C-Arm console.
2.
Release the screws of PCB Box cover, remove Filament Diver Board and insert new one.
3.
Configure the Jumper: J5 for old Monoblock without blue mark or J6 for new Monoblock
with Blue Mark:
The jumper is installed on J6 for new Monoblock as default.
4.
Confirm the switch J7 is set to small position. If not, configure to small focal spot.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...