Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-18
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0205513
Warning
Set Xkey lock
failed
Image processing LEDs on
keyboard display exception.
The display may be abnormal.
Please ensure no key is
pressed on keyboard, then
press “Auto B/C” key and
“Negate” key. If this issue
persists, please call for service.
Check system IO state, and
keyboard connection
0205514
Error
Create thread
failed
Execution thread exception.
Please restart the system. If
this issue persists, please call
for service.
System unknown error
0205515
Error
Start thread
failed
Execution thread exception.
Please restart the system. If
this issue persists, please call
for service.
System unknown error
0205517
Error
Register CORBA
servant failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
Make sure if omniNames
service failed, any module
initialization failed
0205518
Error
Register CORBA
sync notify
servant failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
0205519
Warning
Unregister
CORBA servant
failed
Workstation communication
closing exception. Please
restart the system. If this issue
persists, please call for service.
make sure if omniNames
service failed, check
omniOrb log file
0205520
Error
Subscribe sync
event failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
0205521
Error
Unsubscribe
sync event
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed,
CorbaEventChannel
initialization failed
0205522
Error
Resolve CORBA
object in name
service failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
crashed
0205523
Error
Invoke CORBA
interface failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
crashed
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...